System Specifications
|
Analysis Parameters |
|
PT, APTT, TT, FIB, VT, PC, PS, APC-R, AT-III, D-Dimer |
|
Factor II, Factor V, Factor VII, FactorVIII, Factor IX, Factor X |
|
Factor XI, Factor XII, Heparin |
Principle |
|
Scattered Light Detection Method |
Sample Volume Required |
|
PT: 25ul |
|
APTT: 25ul |
|
TT: 25ul |
|
FIB: 25ul |
Testing Channels |
|
4 Channels |
Sample Positions |
|
24 |
Reagent Positions |
|
6 |
Testing Time |
|
20-90 seconds |
Internal Timer Resolution |
|
< 0.1 s |
Repeatability |
|
CV: PT œ 2%, APTT œ 2%, TT œ 10%, FIB œ 4% |
Linearity |
|
r >= 0.995 |
Temperature Control |
|
Sample Incubator 37.0 + or - 0.2℃ |
|
Reagent Incubator 37.0 + or - 0.2℃ |
Operating conditions |
|
Temperature 10℃-40℃, Humidity 20%-85% |
Display |
|
5.7″ LCD display(320 x 240 resolution, 256 gray scale) |
Input |
|
Touch panel and pen, external mouse |
Data Storage Capacity |
|
10000 test results |
Output |
|
External Printer |
Data Communication |
|
RS-232 serial cable |
Dimensions L × W × H |
|
410 x 320 x 150 |
Weight |
|
10 kg |
Power supply |
|
AC 90V-250V, 50-60Hz |